Hasil (
Bahasa Indonesia) 1:
[Salinan]Disalin!
IntroductionImage Intensifier11-5Figure 1-6. Placement of Collimator and Beam Alignment Test Tools for Checking Fluoro Units When this occurs, the alignment of the center of the x-ray tube to the image intensifier can be checked by noting if the image of the Collimator Tool is symmetrical, as seen on the TV monitor or mirror. While observing the image of the centimeter scales of the Collimator Tool, record the distance to the edge of the viewing field for each of the four axes of the tool. If the x-ray field is centered, these distances should be the same (Figure 1-6). If the centering is off by more than 3% of the SID, it should be corrected by qualified service personnel. Tape a large cardboard cassette, containing a fresh film or a "Ready-Pac", to the input face of the image intensifier. Open the fluoro collimator shutters as far as possible, and expose the cassette in the fluoro mode for about 15 seconds at 60 kVp and 1-2 mA. The spot film device is not used for this test exposure. The processed film will show the image of the Collimator Tool. The extent of the actual field size at the tabletop will be shown as the distance on the Collimator Tool scale that is imaged for each axis of the tool. The difference between the viewed field size and the actual field shown on the film is a collimation error. This test is illustrated in Figure 1-7. Nuclear Associates 07-661-7662 Operators Manual 1-6 Figure 1-7 Test of Viewed Field Versus Actual Field Size This error for any given field edge must be less than 3% of the SID. If the x-ray tube is 45 cm (18”) below the table and the image intensifier is 30 cm (12”) above the table, the allowed error is 2.25 cm (0.9”) in any one direction. 1.6 Checking Spot Film Collimation The spot film collimation may be checked in a similar manner. With the Beam Alignment and Collimator Tools on the table, tape a loaded cardboard cassette (that is larger than the spot film cassettes size) onto the face of the spot film device or grid. Record the size of the spot film cassette and make a full spot film exposure. A loaded cassette can be placed in the spot film device for this test. If the edges of the x-ray field are visible in the spot film, the collimation is good, and film centering with respect to the x-ray field can be checked. If the edges of the x-ray field are not visible on the spot film, the field size shown on the larger film from the cardboard cassette will indicate qualitatively the size of the actual x-ray field. References: Regulations for the Administration and Enforcement of the Radiation Control Health and Safety Act of 1968. U.S. Department of Health, Education, and Welfare. Food and Drug Administration, July, 1980.NOTE(Blank Page)
Sedang diterjemahkan, harap tunggu..
